zum Inhalt


Thomas Kalinowski

Reducing the number of monitor units in multileaf collimator field segmentation

University of Rostock, Institute of Mathematics, 2004

https://doi.org/10.18453/rosdok_id00005902

Abstract: Multileaf collimators (MLCs) are the prevailing tool for the realization of radiation fields in intensity modulated radiation therapy (IMRT). One step in the treatment planning is to determine a set of leaf positions realizing a certain intensity modulated radiation field. In this paper we suggest two extensions in the use of the MLC that lead to considerable savings in terms of monitor units, thus potentially increasing the treatment quality. We test our method with random and with clinical sample matrices.

Workingpaper   Freier Zugang    


Portale

OPACGVKDataCite Commons

Rechte

alle Rechte vorbehalten

Das Werk darf ausschließlich nach den vom deutschen Urheberrechtsgesetz festgelegten Bedingungen genutzt werden.